E – Coil Tester
DPN 402197 © TSS (International) Ltd Page 5 of 10
E.2 OPERATION
The Coil Tester is supplied with default settings of 25Hz. To change the frequency
specified, see Section E.2.1.
Operation of the Coil Tester is very simple procedure outlined in the following steps:
1. Power on the 350 System and ensure the Surface Display Computer (SDC) is
connected to DeepView to confirm the test results. Ensure the coil calibration con-
stants have been entered correctly and the 350 System is setup correctly (see
section E.1.1).
2. Select the Terminal Window into view and set focus to this window. Press
[CTRL+U]. This will display a warning message stating that continuing will stop
communication with the SEP. Click ‘YES’ and the 350 Terminal Menu will be dis-
played.
3. Press [2] to access the Pre-amp Gain Menu. There are two fields on this screen
outlined in Table E–1 below.
To cycle through the available options press [SPACE] bar. Press [ENTER] to accept
settings.
4. As outlined above, set the Pre-amp Gain to 4, Pre-amp Autogain to OFF.
5. When the pre-amp gain and autogain have been set, a chevron will be displayed
on screen.
6. Press [4] to enter the Data Window/FFT Size Menu. Set the tone frequency to
25Hz and press [ENTER] to accept the selection and return to the Terminal Menu.
7. Press [9] to exit the Terminal menu.
8. To re-establish communication with the 350 System, press the Enable/disable
polling on the Terminal Window toolbar. Scrolling data will be displayed on screen
to identify the system is operating correctly.
9. Place the Coil Tester over the end of the search coil so the end of the coil is
seated fully into the circular recess of the tester.
Table E–1: 350 System Subsea Parameters
Parameter Options Required Setting
Pre-amp Gain 1,2,3 and 4 4
Pre-amp Autogain ON or OFF OFF